Compliance information
Compliance information
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument
complies.
EMC compliance
EU EMC Directive
Meets intent of Directive 2014/30/EU for Electromagnetic Compatibility. Compliance was demonstrated to the following
speci
fi
cations as listed in the Of
fi
cial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006.
EMC requirements for electrical equipment for measurement, control, and
laboratory use.
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic
fi
eld immunity
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
IEC 61000-4-11:2004. Voltage dips and interruptions immunity
EN 61000-3-2:2006.
AC power line harmonic emissions
EN 61000-3-3:1995.
Voltage changes,
fl
uctuations, and
fl
icker
European contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
For compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Oscilloscope:
≤
3.0 division waveform displacement and
≤
6.0 division increase in peak-to-peak noise. RF: Residual spurious
signals in the RF section can typically increase to –50 dBm when the instrument is subjected to electromagnetic interference per the
IEC 61000-4-3 test for frequencies up to 1 GHz, and to –35 dBm for frequencies above 1 GHz.
5
Oscilloscope:
≤
1.0 division waveform displacement and
≤
2.0 division increase in peak-to-peak noise. RF: Residual spurious
signals in the RF section can typically increase to –85 dBm when the instrument is subjected to electromagnetic interference
per the IEC 61000-4-6 test.
6
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels (IEC
61000-4-11).
x
MDO3000 Series Oscilloscopes User Manual
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...