Display Waveform or Trace Data
10.
Select
Probe Setup
to de
fi
ne probe
parameters.
On the side menu:
Select
Voltage
or
Current
to set the
probe type for probes that do not
have a TekProbe Level 1, TekProbe II
(requires a TPA-BNC adapter) or TekVPI
interface.
For probes that do not have a Tek
interface, when
Probe Type
is set to
Voltage
, turn
Multipurpose a
to set
Attenuation
to match the probe
For probes that do not have a Tek
interface, when
Probe Type
is set to
Current
, turn
Multipurpose a
to set the
Amps/Volts ratio (Attenuation) to match
the probe.
If you are measuring current by probing
the voltage drop across a resistor,
set
Measure Current
to
Yes
. Push
A/V
ratio on the side menu and turn
Multipurpose a
to set the Amps/Volts
or Volts/Amp ratio of your setup. For
example, if you are measuring the drop
across a 2
Ω
resistor, set the V/A ratio
to 2.
For some types of probes, you can push
this button to instruct the oscilloscope to
perform an AC calibration on the entire
signal path from the probe tip to the
speci
fi
c oscilloscope channel. This can
result in a
fl
atter frequency response
over the entire frequency range.
11.
Select
Deskew
to make display and
measurement adjustments for probes that
have differing propagation delays. This is
especially important when using a current
probe in conjunction with a voltage probe.
For best results, use a deskew
fi
xture, such
as the Tektronix 067-1686-xx.
MDO3000 Series Oscilloscopes User Manual
111
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...