Important safety information
Keep
fi
ngers behind the
fi
nger guards on the probes.
Remove all probes, test leads and accessories that are not in use.
Use only correct Measurement Category (CAT), voltage, temperature, altitude, and amperage rated probes, test leads,
and adapters for any measurement.
Beware of high voltages.
Understand the voltage ratings for the probe you are using and do not exceed those ratings.
Two ratings are important to know and understand:
The maximum measurement voltage from the probe tip to the probe reference lead.
The maximum
fl
oating voltage from the probe reference lead to earth ground
These two voltage ratings depend on the probe and your application. Refer to the Speci
fi
cations section of the manual
for more information.
WARNING.
To prevent electrical shock, do not exceed the maximum measurement or maximum
fl
oating voltage for the
oscilloscope input BNC connector, probe tip, or probe reference lead.
Connect and disconnect properly.
Connect the probe output to the measurement product before connecting the
probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe input.
Disconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe from
the measurement product.
Connect and disconnect properly.
De-energize the circuit under test before connecting or disconnecting the current
probe.
Connect the probe reference lead to earth ground only.
Do not connect a current probe to any wire that carries voltages above the current probe voltage rating.
Inspect the probe and accessories.
Before each use, inspect probe and accessories for damage (cuts, tears, or
defects in the probe body, accessories, or cable jacket). Do not use if damaged.
Ground-referenced oscilloscope use.
Do not
fl
oat the reference lead of this probe when using with ground-referenced
oscilloscopes. The reference lead must be connected to earth potential (0 V).
MDO3000 Series Oscilloscopes User Manual
vii
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...