Use the Arbitrary Function Generator
Quick Tips
You can save arbitrary waveforms in .CSV format. A .CSV
fi
le consists of sets of waveform point pairs (voltage, point
number).
You can load arbitrary waveforms from a variety of active time domain waveform sources: Channel 1 - 4, Ref 1 - 4,
MATH, digital channels D0 - D15.
The arbitrary waveform memory may be saved to and loaded from one of four arbitrary waveform memory locations.
These may only be loaded to the active arbitrary waveform. You cannot save them to
fi
les or recall them directly to
the arbitrary waveform memory locations.
NOTE.
Loading from digital waveform sources will use the main waveform record. MagniVu records are not supported.
NOTE.
Load impedance scales the vertical settings to show what the load would see based on the load impedance. Since
the AFG is a 50
Ω
source, for most accurate operation, set the load impedance to 50
Ω
and set the input channel to 50
Ω
.
NOTE.
Noise range is reduced above 50% of the function Maximum Amplitude according to the function:
Noise Percent Maximum = 100.0 * (Maximum Amplitude / Amplitude – 1.0)
190
MDO3000 Series Oscilloscopes User Manual
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...