Input
Description
RF input maximum input voltage
Average Continuous Power: +20 dBm (0.1 W)
DC maximum before damage: ±40 VDC
Max no 33 dBm (2 W) CW
Peak Pulse Power: +45 dBm (32 W)
Peak Pulse Power defined as <10 μs pulse width, <1% duty cycle, and reference
level of ≥ +10 dBm
Table 13: Maximum input with a P6316 Digital Probe
Input
Description
Threshold Accuracy
±(100 mV + 3% of threshold setting after calibration)
Threshold Range
+25 V to –15 V
Maximum nondestructive input signal to probe
+30 V to -20 V
Minimum signal swing
500 mV
peak-to-peak
Input resistance
101 kΩ
Input capacitance
8.0 pF typical
Pollution Degree
2, indoor use only
Humidity
5% to 95% relative humidity
Powering the oscilloscope
Use this procedure to connect the oscilloscope to line power and power on and off the oscilloscope. Always connect the oscilloscope to AC
power using the power cord that shipped with the instrument.
Before you begin
Prerequisite: Use the AC power cord that shipped with your oscilloscope.
Procedure
1. Connect the supplied power cord to the oscilloscope power connector.
Install your instrument