DXF-200 Getting Started Guide
Page 26
Connecting the Utilities
Connecting the Power
Plug the
SYSTEM POWER
cord on the back of the DXF to an electrical outlet rated in accordance with
the power input information on the back of the instrument.
CAUTION: Whenever plugging or unplugging power cords, handle them by the plugs, not by the
cords.
MISE EN GARDE: Chaque fois que vous branchez ou débranchez les cordons d'alimentation,
tenez-les par les fiches et non par les cordons.
CAUTION: An independent heavy GROUND wire must be provided through the power hook up.
Improper grounding may cause severe damage for which the supplier will not accept responsibility.
All power strips must be fully grounded and carry the ground through to the sockets into which the
computer is plugged.
MISE EN GARDE: Un fil de masse lourd et indépendant doit être fourni via le raccordement élec-
trique. Une mise à la terre incorrecte peut causer de graves dommages pour lesquels le fournisseur
décline toute responsabilité. Toutes les barrettes d'alimentation doivent être entièrement mises à la
terre et acheminer la terre jusqu'aux prises dans lesquelles l'ordinateur est branché.
Connecting the LN2 Lines
1
Obtain an LN2 tank that can provide a maximum of 22 psi per the Thermal Diffusivity DXF 200 Site
Preparation Guide.
2
Connect the
LN2 LIQUID OUTLET
on the tank to the
LN2 INLET
fitting on the right side of the EM.
See
.
3
Connect the
CRYOGENIC VALVE
cable to the
LN2 OUTLET
).
Figure 10
LN2 tank connections.
CRYOGENIC VALVE
LN2 LIQUID OUTLET