DXF-200 Getting Started Guide
Page 12
Table 2:
DXF Front Panel LED Indicators - Temperature Displays
Indicator
Function
FURNACE TEMPERATURE°C
Indicates the current temperature of the furnace. There are
several indicators just to the left of the temperature display
that may light up during normal operation.
SPECIMEN TEMPERATURE °C
Indicates the current temperature of the sample. There are
several indicators just to the left of the temperature display
that may light up during normal operation.
OVER TEMP
When the temperature controllers detect a temperature out-
side the acceptable range, an alarm is triggered, and the
OVER TEMP red LED will light up. If the interlock ON LED
is enabled, it will be disabled.
OPTICAL
When the internal interlocks for the optical path are enabled
and the operating software enables it, the OPTICAL LED
lights up. When the OPTICAL LED is ON, the flashlamp is
enabled to fire.
ALARM
This indicator shows the status of the External Alarm input.
This unit does not have an External Alarm Input and this
LED will not be on.
POWER
When the internal 12 VDC power supply is ON, the yellow
POWER LED lights up.
DET. POWER
When the operating software is ready to start a test, a separate
DC power supply is connected to the Detector Electronics
and the DET POWER LED lights up.