DXF-200 Getting Started Guide
Page 24
DXF Top Panel
The DXF top panel has three ports; the table below provides a description of the function of each port.
Refer to Figure 7 for an illustration of rear connections.
Figure 9
Environmental Module top panel connections.
Connecting the Environmental Module to the Pulse Source
To connect the DXF to the EM, follow the instructions below. Refer to
above for
connection ports, or the diagram located in the
1
Connect the
SOLID STATE DETECTOR
cable from the back of the head plate on the EM to the
J102
port on the back of the DXF.
2
Connect the
TEMPERATURE SENSORS
cable from the side of the EM to the
port on the top
of the DXF.
3
Connect the
HEATER POWER
cable from the back of the EM to the
port on top of the DXF.
4
Connect the
CRYGENIC VALVE CONNECTION
from the
LN2 OUTLET
on the tank to the
port on top of the DXF.
Port
Function
TEMPERATURE SENSORS
Connect Furnace and Sample RTD wires to Controls
HEATER POWER
Connect Heater Wires to Controls
CRYOGENIC VALVE CONNECTION
Connect the Cryogenic Cooling Valve to the Controls
TEMPERATURE
SENSORS (Sample and
Control)
HEATER POWER
SOLID STATE
DETECTOR cable to the
DXF
CRYOGENIC VALVE
CONNECTION
2
3
1
2
PURGE GAS tubing
connected to the PURGE
GAS INLET
3
2
1