INSTRUCTION MANUAL
66
8.6 SETTING OF THE EMISSIVITY
1. The LC-display as well as the push buttons for displaying and setting of the parameters
are found inside the unit. The pyrometer is opened by 4 allen screws (1). If unscrewed,
the rear cover can be pulled out along with the attached display and push buttons. The
pullout is limited by the lengths of the screws. The backlight of the display is always
powered in either status, opened or closed pyrometer. Open the four screws from the
back of the camera and pull the card outwards.
1
2
3
4
1. Screws of the card, 2. Pulling direction, 3. PAR-button, 4. ESC-button, 5. Laser targeting
light push button
2. Check the correct value for the used material from the table below. Press the “PAR”-
button once to set parameter. Insert the value with arrow keys. Accept the value with
“ENT”- button.
(Pushing the ESC button changes the pyrometer to measuring mode. If a parameter is
changed with the arrow keys the indication of the ESC button changes to ENT. Pressing
the button again confirms the value into the pyrometer. Changing the parameters again
by pushing the PAR button doesn’t confirm this value in the pyrometer. If no button is
pressed for 30 s, the pyrometer changes to the temperature indication without accepting
the changed value).