Air Power Line II Instructions for Use
DePuy Synthes
31
2. Manipulate moving parts
Manipulate all moving parts such as triggers, sliding sleeves,
attachment release rings, saw blade coupling , locking knob
for the saw blade quick coupling and switches under running
tap water to loosen and remove gross debris.
3. Spray and wipe
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes. Follow the enzymatic
detergent manufacturer’s directions for correct temperature,
water quality (i.e. pH, hardness) and concentration/dilution.
4. Clean with detergent
Clean the device manually under running warm water using
an enzymatic cleaner or detergent for a minimum of 5 min-
utes. Manipulate all moving parts under running water. Use
a soft-bristled brush and/or soft lint-free cloth to remove all
visible soil and debris. Follow the enzymatic cleaner or deter-
gent manufacturer’s instructions for use for correct tempera-
ture, water quality and concentration/dilution.
5. Rinse with tap water
Rinse the device thoroughly using cool to lukewarm running
water for a minimum of 2 minutes. Use a syringe or pipette
to flush lumens and channels. Actuate joints, handles and
other movable device features in order to rinse thoroughly
under running water.
6. Visually inspect device
Inspect the cannulations, sliding sleeves, attachment release
rings, etc. for visible soil. Repeat steps 1–6 if visible soil re-
mains.
7. Load the washing tray
Place all articles in the washing tray in a way that an effec-
tive washing/disinfection can be performed. Ensure that the
attachments are positioned in an upright position and fully
opened. Ensure that the water can flow off any surface.
Damage due to improper reprocessing is not covered by the
warranty.