24
DePuy Synthes
Air Power Line II Instructions for Use
Precautions:
– Reprocessing must be performed immediately after each
use.
– Cannulations, unlocking sleeves and other narrow sites
require special attention during cleaning.
– Cleaners with a pH of 7–9.5 are recommended. The use
of cleaners with higher pH-values can – depending on the
cleaner – cause dissolution of the surface of aluminum,
titanium and its alloys, plastics or compound materials.
The use of such cleaners should be subject to the data
regarding material compatibility in the corresponding data
sheet.
At pH values higher than 11, the surface of stainless steel
can be affected.
For detailed information about material compatibility,
refer to the document “Important Information” at
www.synthes.com/reprocessing. Please refer to the
chapter “Material Compatibility of Synthes Instruments in
Clinical Processing”. Concerning the clinical reprocessing
of the Air Power Line II system please refer to the follow-
ing section of this document.
– Follow the enzymatic cleaner instructions for use for
correct dilution/concentration, temperature and water
quality. Devices should be cleaned in a fresh, newly-made
solution.
– Detergents used on the products will be in contact with
the following materials: stainless steel, aluminum, plastic,
and rubber seals.
– Never immerse the handpiece or attachments in aqueous
solutions or in an ultrasonic bath. Do not use pressurized
water as this will cause damage to the system.
– Synthes recommends using new sterile cutting tools for
each operation. Refer to “Clinical Processing of Cutting
Tools” (036.000.499) for detailed clinical processing in-
structions.
Power tool units and attachments are frequently exposed to
high mechanical loads and shocks during use and should not
be expected to last indefinitely. Proper handling and mainte-
nance help extend the useful life of surgical instruments.
Gentle care and maintenance with proper lubrication can
substantially increase the reliability and life of the system
components and reduce the risk of malfunction or harm to
the user and patient.
Synthes power tools must be serviced and inspected annu-
ally by the original manufacturer or an authorized site. Yearly
maintenance will ensure that the equipment maintains the
highest standard of performance and will prolong the life of
the system. The manu facturer assumes no warranty for dam-
ages arising from improper use, neglected or unauthorized
servicing of the tool.
For more information about Care and Maintenance, please
refer to the Air Power Line II Care and Maintenance Poster
(038.000.017).
General Information
Care and Maintenance