1
2
3
Air Power Line II Instructions for Use
DePuy Synthes
33
Prior to maintenance refer to the following procedures:
Article number
Procedure
Air Power Line II handpieces
Remove the seal nipple (519.596) on the
machine’s air inlet.
511.606
511.610
511.615
Synthes Double Air Hoses
Ensure that the inlet and outlet of
the Synthes Double Air Hose are not
connected.
519.510
519.530
519.550
Dräger Double Air Hoses
Remove the seal nipple (519.596) from the
Dräger Double Air Hose and ensure that
the inlet and outlet are not connected.
519.610
519.630
519.650
BOC/Schrader Double Air Hoses
Remove the seal nipple (519.591 or
519.592) from the BOC/Schrader Double
Air Hose and ensure that the inlet and
outlet are not connected.
519.511
519.531
Oiling the power tools
– After each use, apply around 5 drops of the Synthes
special oil (oil dispenser 519.970) into the air inlet of the
cleaned power tool (Fig. 1).
– Connect the power tool to a double hose using the lubri-
cation adapter (519.790) (Fig. 2).
– Wrap gauze or a cloth around the hose coupling to ab-
sorb the exiting oil.
– Run the power tool for approximately 20 seconds.
– If dirty oil exits, the process must be repeated.
Make sure that the unlocking ring (Page 8
2
), the trigger
and sliding sleeve (Page 18
2
,
3
), the forward and reverse
knob (Page 8
5
) and the saw blade coupling (Page 18
1
,
top arrow) are lubricated after each use. After lubricating the
trigger shaft, press the trigger several times.
Oiling the attachments
After each use, oil the rings on all the cleaned attachments
and the jaws of the cleaned drill chuck with 1 drop of
Synthes special oil (Fig. 3).
Distribute the oil by moving the parts, and remove excess oil
with a cloth.
Maintenance and Lubrication