Air Power Line II Instructions for Use
DePuy Synthes
35
Inspection and Function Test
Visually inspect for damage and wear (e.g. unrecognizable
markings, missing or removed part numbers, corrosion, etc.).
Check the handpiece controls for smooth operation and
function.
All movable parts should be moving smoothly. Check that
the triggers do not remain blocked in the handpiece when
pressing on them. Check that no residuals prevent the mov-
able parts from moving smoothly.
Check the release ring of the handpiece and attachments for
smooth operation, and check for function together with cut-
ting tools.
Check instruments and cuttings tools for correct adjustment
and functioning prior to every use.
Do not use damaged, worn or corroded components but
send them instead to the Synthes Service Center.
Failing to follow these instructions will lead to damage and
malfunction, increasing the risk of harm to the user and pa-
tient. For further information on inspection and function
test, please refer to the APL II Care and Maintenance Poster
(038.000.017).