Air Power Line II Instructions for Use
DePuy Synthes
27
Cleaning brush (516.101)
Manual Cleaning Instructions
1. Remove debris
Rinse the device under running cold tap water for a mini-
mum of 2 minutes. Make sure that no liquids enter the air in-
let. Plug the air inlet with the seal nipple (519.596).
Use a sponge, soft lint-free cloth or soft-bristled brush to as-
sist in removing gross soil. For cannulations of the handpiece
and attachments, the cleaning brush (516.101) shown below
should be used. (Fig. 5)
Precautions:
– Do not use pointed objects for cleaning.
– Brushes and other cleaning tools shall be either single-use
items or, if reusable, be decontaminated at least daily us-
ing a solution as detailed in section “3. Spray and wipe”.
– Brushes shall be inspected before daily use and discarded
if they have degraded to the point where they may
scratch instrument surfaces or be ineffective due to worn
or missing bristles.
– Make sure that no cleaning solution enters the machine’s
air inlet.
– Never immerse the handpiece or attachments in aqueous
solutions or in an ultrasonic bath. Do not use pressurized
water as this will cause damage to the system.
2. Manipulate moving parts
Manipulate all moving parts such as triggers, sliding sleeves,
attachment release rings, saw blade coupling, locking knob
for the saw blade quick coupling and switches under running
tap water to loosen and remove gross debris.
3. Spray and wipe
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes. Follow the enzymatic
detergent manufacturer’s directions for correct temperature,
water quality (i.e. pH, hardness) and concentration/dilution.