Air Power Line II Instructions for Use
DePuy Synthes
41
If the recommended solutions do not work, send the power tool to your local
Synthes Service Center.
For further technical questions or information on our services, please contact
your Synthes representative.
Problem
Possible causes
Solution
Attachments cannot be coupled to
the Air Reamer/Drill.
Attachment coupling and/or unlocking
sleeve are blocked.
Lock the power tool. Remove particles
with a tweezer. Do not use sharp ob-
jects for this. If necessary, clean and oil
the power tool and move the moving
parts a few times.
Tool cannot be coupled.
Coupling geometry of the tool has
changed due to wear.
Exchange the tool, or send it to your
local Synthes Service Center.
Bone and tool heat up during surgery.
The tool blades are blunt.
Replace the tool.
Oscillating saw vibrates too much.
The saw blade has come loose.
Tighten the fixation knob for the saw
blade quick coupling more firmly.
Saw blade is difficult to couple or
cannot be coupled.
General wear and tear has affected
the connection geometry of the saw
blade.
Replace the saw blade.