78
A-96.250.501 / 280915
AMI Hydrazine
Default Values
10. Default Values
Operation:
Sensors:
Filter Time Const.:..................................................................30 Sec
Hold after Cal.:.....................................................................300 Sec
Alarm Relay
...................................................................... same as in Installation
Relay 1 and 2
...................................................................... same as in Installation
Input
...................................................................... same as in Installation
Logger:
Logger Interval:.............................................................. 30 Minutes
Clear Logger: ................................................................................no
Installation:
Sensors
Temp. comp.: .............................................................................. 3%
Signal Output 1
Parameter: ........................................................................Hydrazine
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ............................................................0.00 ppb
Scaling: Range high:.............................................................100ppb
Signal Output 2
Parameter: ....................................................................Temperature
Current loop: .....................................................................4 –20 mA
Function: .................................................................................. linear
Scaling: Range low: ................................................................0.0 °C
Scaling: Range high:.............................................................50.0 °C
Alarm Relay:
Alarm Hydrazine:
Alarm high:..........................................................................1000ppb
Alarm low: ...........................................................................0.00 ppb
Hysteresis: ..............................................................................10ppb
Delay:.......................................................................................5 Sec
Sample Flow; Flow Alarm ........................................................... yes
Sample Flow; Alarm high.....................................................300 rpm
Sample Flow; Alarm low ......................................................150 rpm
Sample Temp; Alarm high:.......................................................55 °C
Sample Temp; Alarm low: ..........................................................5 °C
Case temp. high:......................................................................65 °C
Case temp. low: .........................................................................0 °C
Relay 1 and 2
Function: .......................................................................... limit upper
Parameter: ........................................................................Hydrazine
Setpoint:.............................................................................1000 ppb
Hysteresis: .............................................................................10 ppb
Delay:.....................................................................................30 Sec