16
A-96.250.501 / 280915
AMI Hydrazine
Installation
3.4.
Install the Reference Electrode
The reference electrode is delivered separately and protected with
a waterfilled protective cap. The connector is fixed to the panel with
an adhesive tape and already connected to the front end PCB in
the AMI transmitter.
To install the reference electrode proceed as follows:
1
Loosen the union nut [D].
2
Remove the protective cap [B] from the reference electrode [C].
3
Push the reference electrode through the union nut [D] into the
bore of the flow cell block [E] as far as it will go.
4
Tighten the union nut.
5
Remove the connector [A] from the panel and screw it onto the
reference electrode.
A
B
C
Connector
Protective cap
Reference sensor
D
E
Union nut
Flow cell block
C
B
D
E
A