AMI Hydrazine
Program List and Explanations
A-96.250.501 / 280915
71
5.3.2 and 5.3.3
Relay 1 and 2:
The contacts can be set as normally open or nor-
mally closed with a jumper. See
. The
function of relay contacts 1 or 2 are defined by the user.
Note:
The navigation in the menu <Relay 1> and <Relay 2> is
identical. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1
First select the functions as:
- Limit upper/lower
- Control upwards/downwards
- Timer
- Fieldbus
2
Then enter the necessary data depending on the selected func-
tion.
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20
Parameter
: select a process value
5.3.2.300
Setpoint
: If the measured value rises above respectively falls below
the set-point, the relay is activated.
5.3.2.400
Hysteresis:
within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
5.3.2.50
Delay:
Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range. 0–600 Sec
5.3.2.1
Function = Limit upper/lower:
Parameter
Range
Hydrazine
0.00–1000 ppb
Temperature
-30 to + 120 °C
Sample flow
0–300 rpm
Parameter
Range
Hydrazine
0.00–1000 ppb
Temperature
0 to + 100 °C
Sample flow
0–300 rpm