AMI Hydrazine
Troubleshooting
A-96.250.501 / 280915
51
7.
Troubleshooting
This chapter provides some hints to make troubleshooting easier.
For any detailed information how to handle or clean parts please see
. For any detailed information how to program the
instrument please see
Program List and Explanations, p. 61
.
Process
calibration
Possible error message
Current value too low or no sample flow.
Offset Error
Slope Error
Current value too low or no sample flow:
Offset Error:
Possible cause
Corrective action
The ppb value of the sample is
too low, thereby, the current dif-
ference to zero is too small.
Process value used for
calibration should be higher.
Check the diagnostic value of
the zero point.
Low sample flow resulting in no
signal difference to zero.
Check sample inlet pressure, if
necessary readjust sample flow
to min. 170 rpm.
Hydrazine sensor contaminated
Sensor does not provide enough
current for the value measured
with the manual analysis.
Clean sensor, see
In case of repeated sensor con-
tamination, check for water treat-
ment chemicals such as
phosphates.
Possible cause
Corrective action
Run-in time of sensor to short.
Let the sensor run-in at least 30
min
Zero point calibration was too
short.
Repeat the zero point calibration