22
A-96.250.501 / 280915
AMI Hydrazine
Installation
3.7.
Relay Contacts
Programming of the relay contacts see
3.7.1
Input
Note:
Use only potential-free (dry) contacts.
• Terminals 16/42
If signal output is set to hold, measurement is interrupted if input is
active.
For programming see menu
3.7.2
Alarm Relay
Note:
Max. load 1 AT / 250 VAC
Alarm output for system errors.
Error codes see
NOTICE:
With certain alarms and certain settings of the AMI
transmitter the alarm relay does not switch. The error, however,
is shown on the display.
1)
usual use
Terminals
Description
Relay connection
NC
1)
Normally
Closed
10/11
Active (opened) during normal
operation.
Inactive (closed) on error and
loss of power.
NO
Normally
Open
12/11
Active (closed) during normal
operation.
Inactive (opened) on error and
loss of power.
10
12
11
0V
1)
10
12
11
0V