AMI Hydrazine
Program List and Explanations
A-96.250.501 / 280915
65
5 Installation
5.1 Sensors
5.1.1
Temp Comp
Temperature compensation in %.
Range: 0.0–5.0%
5.2 Signal Outputs
Note:
The navigation in the menu <Signal Output 1> and
<Signal Output 2> is identical. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
5.2.1 and 5.2.2
Signal Output 1 and 2:
Assign process value, the current loop
range and a function to each signal output.
5.2.1.1
Parameter:
Assign one of the process values to the signal output.
Available values:
Hydrazine
Temperature
Sample Flow
5.2.1.2
Current Loop:
Select the current range of the signal output. Make
sure the connected device works with the same current range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3
Function:
Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See
Control upwards or control downwards for controllers.
See