AMI Hydrazine
Instrument Setup
4.
Instrument Setup
4.1.
Install the Diisopropylamine Bottle
Only install the diisopropylamine bottle if you intend to start opera-
tion immediately. Do not it install if no sample in available!
Fill the Diisopropylamine into the glass bottle (G45 thread), deliv-
ered with the instrument. If you want to connect a Merck bottle di-
rectly, you need a thread adapter which can be ordered at SWAN,
order number C-83.591.010.
WARNING
Diisopropylamine is corrosive.
Read the Safety Data Sheets (SDS) first.
Wear suitable protective clothing, gloves and eye/face pro-
tection.
Avoid inhalation of DIPA vapor. To prevent formation of re-
agent vapors:
– close the reagent bottle firmly
– check the EPDM seal regularly
In case of contact with eyes, rinse immediately with plenty of
water eyelid wide open for at least 10 min, summon medical
advice. In case of accident or if you feel unwell, summon
medical advice immediately (show the label where possible).
A
B
C
D
E
F
Bottle holder
Screw cover G 45
EPDM seal mounted on the bottle holder
Reagent bottle
Steel tube
Diffusions tube