36
A-96.250.501 / 280915
AMI Hydrazine
Maintenance
6.2.
Stop of Operation for Maintenance
WARNING
Electrical shock hazard!
Do not carry out maintenance work during normal operation.
Always turn off power before manipulating electric parts.
WARNING
Diisopropylamine is corrosive.
Read the Safety Data Sheets (SDS) first.
Wear suitable protective clothing, gloves and eye/face pro-
tection.
Avoid inhalation of DIPA vapor.
In case of contact with eyes, rinse immediately with plenty of
water eyelid wide open for at least 10 min, summon medical
advice. In case of accident or if you feel unwell, summon
medical advice immediately (show the label where possible).
1
Wear safety gloves and safety glasses!
2
Remove the diisopropylamine bottle and close it well.
3
Screw an empty bottle on the holder. Let the sample run for 30
min.
4
Stop sample flow by closing the flow regulating valve.
5
Wait until the rotor stops and hydrazine reading is 0 ppb.
6
Shut off power of the instrument.
7
Empty the constant head by opening the grab sample tap.