SVAN 106 User MANUAL
_
25
7. Press the <ENTER> push-button in order to accept the measurement
result.
The calibration factor is calculated, stored and displayed after pressing the
<ENTER> push-button.
4.3.5.
History of the calibrations – Calibration History
The Calibration History window displays
up to ten last calibration records.
<ENT>
In order to review all calibration record, the user has to use the <
5
5
5
5
>, <
6
6
6
6
>
push-buttons. The opened window contains the date and time of the
performed calibration measurement, the way the calibration was done
(Calibr. By Measurement or Calibr. By Sensitivity) and the obtained
calibration factor (Calibration Factor).
In the case when the calibration measurements were not performed, the
Calibration History window does not contain any record. The content of
this window is cleared after the Clear Calibr. History operation.
4.3.6.
Clear calibration records - Clear Calibr. History