SVAN 106 User MANUAL                                                                                                                    

                                                                                                         _ 

40 

 

in  the  case  of  Level –  is  selected,  the  triggering  condition  is  fulfilled  only  when  the  Source  has  the 
lower value than this determined in the Level position and in the other cases the triggering condition is 
not fulfilled. 

 

in the case of Slope + is selected, the triggering condition is fulfilled only when the arising value of the 
Source is passing the level determined by the Level parameter.  

 

in the case of Slope – is selected, the triggering condition is fulfilled only when the falling down value 
of the Source is passing the level determined by the Level parameter.   

 

in the case of Gr is selected, the triggering condition is fulfilled only when the signal has the 
greater  level  than  this  determined  in  the  Level  and  the  gradient  of  the  signal  is  greater  than  this 
determined in the Gradient  position. In the other cases the triggering condition is not fulfilled. 

 

in the case  of Gradient - is selected, the triggering condition  is fulfilled  only  when the signal has the 
lower  level  than  this  determined  in  the  Level  and  the  gradient  of  the  signal  is  lower  than  this 
determined in the Gradient  position. In the other cases the triggering condition is not fulfilled. 

Step 

for 

checking 

the 

triggering 

condition  

The  Trig. Step  position  enables  one  to 
select 

integration 

time 

for 

condition 

evaluation:  Logger step,  100ms,  1.0s
Meas. Time and Meas. Period.  

 

 <

4

4

4

4

 

 

Source  for  triggering  condition  and 
channel of triggering signal
 

The  Source  position  enables  one  to  select 
the  type  of  source  for  triggering  condition 
calculation: Vector or Profile 1

The Channel position enables one to select 
the channel of triggering source. Depending 
on  value  of  the  Source  parameter,  the 
value of Channel will be different. 

 

 <

4

4

4

4

 

 

Function for triggering condition definition and threshold 

The  Result  position  enables  one  to  select  the  function  for  triggering 
condition:  PEAK,  P-P,  MAX,  MIN,  RMS  or  VDV. When  Vector  is  selected 
as a Source the only one function is available – RMS

The  Level  position  enables  one  to  select  the  value  of  threshold  for 
triggering condition. The level of the triggering source can be set in a range 
from  60 dB  to  200 dB  or  from  1.00 mm/s

2

 

to  10.0 km/s

2

,  depending  on 

what scale type was selected in the Scale position (path: <Menu> / Display 
/ Results Scale
).  

Speed of the triggering signal change 

This position appears when the Gradient - or Gr trigger is chosen. 
The speed of the triggering signal changes (Gradient) can be set from 1 dB 
to 100 dB range. Speed is defined as dB per Logger Step

 

 

 

Содержание SVAN 106

Страница 1: ...SVAN 106 User MANUAL _ 1 SVAN Vibration Meter USER S MANUAL SVANTEK Sp z o o WARSAW August 2012...

Страница 2: ...ents the software revision named 3 13 1 cf the description of the Unit Label position of the Instrument list The succeeding software revisions marked with the bigger numbers can slightly change the vi...

Страница 3: ...sducer s sensitivity Calibr By Sensitivity 23 4 3 4 Calibration by measurement Calibr By Measurement 23 4 3 5 History of the calibrations Calibration History 25 4 3 6 Clear calibration records Clear C...

Страница 4: ...ting new catalogue and new file 64 7 2 3 Deleting all files from Internal memory Delete All 64 7 2 4 Merging result and setup files memory Defragmentation 65 7 2 5 Opening file catalogue Open 66 7 2 6...

Страница 5: ...meters of 1 1 Octave and 1 3 Octave analysis 83 10 3 Activation of saving of 1 1 Octave and 1 3 Octave analysis results in the logger s file Logger Results 84 10 4 Selection of the 1 1 Octave and 1 3...

Страница 6: ...itionally for measurements of very high impulse vibration the special adapter SA 55 with low pass mechanical filter protecting accelerometer from DC shift effect is available Evaluation of the grip fo...

Страница 7: ...SB 1 1 cable typical 2m SC 118 LEMO 4 pin to LEMO 5 pin connector SV 38V triaxal accelerometer for Whole Body measurements MEMS type SV 105 integrated triaxal Hand Arm adapter with grip force sensor u...

Страница 8: ...the Alt push button Shift The second function of a push button written in red colour on a push button can be used when the Shift push button is pressed This push button can be used in two different w...

Страница 9: ...er access to the frequently used sub lists Save This push button pressed together with the Alt one enables the user to save measurement results as a file in the internal instrument s memory or on the...

Страница 10: ...ode of result s presentation programme the Real Clock RTC and Timer Info The Info push button simultaneous pressing the 3 3 3 3 4 4 4 4 push buttons opens the window with the help information in the m...

Страница 11: ...plot with markers the user has to transfer data to the proper software 20 30 40 50 60 70 80 13 30 00 13 30 09 13 30 17 13 30 26 13 30 35 13 30 43 13 30 52 Leq Marker 1 Marker 2 Marker 3 Marker 4 2 2 I...

Страница 12: ...additional multi purpose input output socket called I O is a two pins LEMO socket On this socket in the case when the Analogue Output functionality is selected the signal from the input of the analogu...

Страница 13: ...nd list Main menu The main list contains the headers of six lists which also contain sub lists or positions The main list is opened after pressing the Menu push button This list contains the following...

Страница 14: ...of different options from which the only one may be selected The selection of the option is performed next way The user has to highlight the desired option by means of 5 5 5 5 or 6 6 6 6 push buttons...

Страница 15: ...der for the printed reports from the measurements This window consists the help information to guide the user how to edit the text The displayed inversely character may be edited One can select the po...

Страница 16: ...g sources Four AA standard internal batteries In the case of alkaline type fully charged set can operate more than 12 h 6 0 V 1 6 Ah Instead of the ordinary four AA rechargeable batteries can be used...

Страница 17: ...s the Alt and Start Stop push buttons in parallel The instrument passes the self test after switching on in this time the producer and the name of the instrument is displayed on the display and then i...

Страница 18: ...n the top of the display The type of measurement function and the measurement mode LM DLM 1 1 and 1 3 etc as well as RTC is also displayed in the same line together with icons The meanings of the icon...

Страница 19: ...s Icon corresponds to the batteries state three two one or none vertical bars in side of the icon When voltage of batteries is too low the icon became red Shift icon is displayed when the Shift push b...

Страница 20: ...press the ENTER push button After the new window is opening one should select the Set Working Directory position and press the ENTER push button again The icon on the upper line and the directory pat...

Страница 21: ...functions In all functions it is possible to perform additionally dose measurements Dose parameters are setting up in HAV WBV Dose window path Menu Measurement In order to select the required function...

Страница 22: ...ition enables the user to Download TEDS Data when accelerometer is connected during instrument s working session It enables also to Upload TEDS Data from the instrument to the accelerometer s TEDS mem...

Страница 23: ...sitivity of the accelerometer taken from its calibration certificate using the 3 3 3 3 4 4 4 4 push buttons or combination of the Shift and 3 3 3 3 4 4 4 4 push buttons The calibration factor is calcu...

Страница 24: ...nt time is also predefined to 5 seconds During the calibration period the ESC and Pause push buttons do not operate but it is possible to stop the measurement using the Start Stop push button Waiting...

Страница 25: ...calibration record the user has to use the 5 5 5 5 6 6 6 6 push buttons The opened window contains the date and time of the performed calibration measurement the way the calibration was done Calibr By...

Страница 26: ...in the Calibration sub list and press ENTER The instrument requests the confirmation of the operation The next pressing of the ENTER push button when the No option is selected causes the closing of th...

Страница 27: ...he Measurement menu can be changed only when the instrument does not execute a measurement The parameters are displayed with different colour and any marker movement is impossible The blinking play ic...

Страница 28: ...in the range of from 1 s to 59 s with 1 second or 10 seconds step from 1 m min to 59 m with 1 minute or 10 minutes step from 1 h to 24 h with 1 hour or 10 hours step It is also possible to set Inf val...

Страница 29: ...easurement ENT Dosimeter parameters can be setup in the HAV WBV Dosimeter window opened from the Measurement list If the Dosimeter function is switched on then the HAV WBV Dosimeter position in the Me...

Страница 30: ...etup Instrument Mode If the Event Recording function is switched off then the Event Recording position in the Data Logging list will be not active 5 2 Setting the parameters for dose measurements HAV...

Страница 31: ...dose measurements The Standard position enables the user to set the standards for the measurements of the HAV WBV Dosimeter The available values of this position are U K Italy Poland France Germany a...

Страница 32: ...channels will be disabled and will not be displayed Activation of second profiles In the Enabled 2nd profiles the user may switch on or off result calculations for the second profiles If second profil...

Страница 33: ...always available filter HP Second available filter for profile 2 is one of BL Wh BL Wk BL Wd BL Wc BL Wj BL Wm BL Wg BL Wb and BL Wf according to the rule if Wh filter is selected in the profile 1 th...

Страница 34: ...lculate vector with other than standard coefficients it is possible to select the coefficient within the values from 0 00 to 2 00 The values presented above are taken into account during the calculati...

Страница 35: ...nnels and Wave Trigger 5 6 The results history logging When Logger function is chosen for the Logging Mode the Data Logging list enables one to program saving the history of results in the logger file...

Страница 36: ...tion of the user any push button should be pressed except the Shift or the Alt one The main measurement results cf App B are calculated in the period set on in the Meas Period These results can be sav...

Страница 37: ...od from N 1 T to NT sig1 svn sig2 svn sigN 1 svn sigN svn time time history PEAK MAX MIN or RMS LOGGER STEP 0 NT LOGGER ON RMS results logged with LOGGER STEP Number of results equal NT LS logger1 svn...

Страница 38: ...s with names Channels x y Channels x y Profile z Channel x Channel x Profile y The VDV history will not be recorded if Wh filer is chosen in this profile Activation deactivation can be done by means o...

Страница 39: ...Measurement Data Logging Logger Setup all measurement results up to the moment the triggering condition disappears the results registered directly after the moment the triggering condition disappears...

Страница 40: ...iggering condition is not fulfilled Step for checking the triggering condition The Trig Step position enables one to select integration time for condition evaluation Logger step 100ms 1 0s Meas Time a...

Страница 41: ...position is selected then event recording will start by trigger Trigger condition is set up in the window opened by position Trigger Parameters This position appears in the list after activation the T...

Страница 42: ...d When Max Length is chosen then the signal will be recording during period defined in the Rec Time but can be stopped earlier if trigger condition was ceased When Fixed Len is chosen then the signal...

Страница 43: ...tive It only informs about the used type of wave recording Continuous The user may define the format of the wave file header Format Format of the wave file header may be PCM or Extensible The Sampling...

Страница 44: ...next triggering condition appears then the signal will be recording additional Rec Time 4 4 4 4 The position Trigger Parameters enables to define the parameters of the triggering signal To open this...

Страница 45: ...control code remotely The next sources of the triggering signal are available RMS 1 VEC46 VEC13 or External External source means that the triggering will take place from the positive or negative slop...

Страница 46: ...ion path Menu Display Results Scale Setting the speed of the triggering signal changes This position appears when the Gradient or Gradient trigger is chosen The speed of the triggering signal changes...

Страница 47: ...ut ENT The Alarm Trigger window and meaning of all positions is identical as for Logger Trigger case 5 11 Programming the instrument s internal timer Timer The Timer position appears only in Advanced...

Страница 48: ...mer can be programmed up to one month ahead and during the date setting the current state of the Real Clock is taken into account The required date can be selected in the special window which is opene...

Страница 49: ...SVAN 106 User MANUAL _ 49...

Страница 50: ...presentation Results Scale enables one to change the scale of result s presentation Screen enables one to set the brightness and the switch on off the screen saver of the instrument s display 6 1 Sel...

Страница 51: ...ode 6 Channels presentation mode 6 Profiles presentation mode When all display modes in the Display Modes list are switched on they all are available and can be selected by means of Alt and 5 5 5 5 6...

Страница 52: ...shows the current second of the measurement The value presented there belongs to the range 1 Meas Period 5 The value of measured function 6 Units of measured value Changing the active fields The chang...

Страница 53: ...urement Channels Setup the 6 Channels mode is not active 1 Result line for Channel 1 2 Result line for Channel 2 3 Result line for Channel 6 4 Function name RMS VDV CRF OVL TIME PEAK P P and MTVV 5 Th...

Страница 54: ...usly results for three profiles 1 Line for first result 2 Line for second result 3 Line for third result 4 Function name RMS VDV CRF OVL TIME PEAK P P and MTVV 5 File name when Auto Save function is a...

Страница 55: ...gger mode can be activated or deactivated in the Display Modes window 1 Logger plot 2 Result value for the cursor position 3 Name of the logged result 4 Name of the logger file 5 Cursor 6 Cursor posit...

Страница 56: ...vailable Linear and Logarithm In the case of the first one the graphical presentation and the units both are linear In the latter case the graphical presentation is given in the logarithmic scale and...

Страница 57: ...Y This position is under development 6 3 Setting the scale of result s presentation Results Scale The Results Scale sub list enables the user to change the scale in the available modes of the measurem...

Страница 58: ...saver function The saving of the internal source of the instrument s power can be achieved by means of reducing the brightness of the screen when possible It is possible to set the Power Saver automat...

Страница 59: ...logger file stored in the wave file settings Result files can be saved manually or automatically Setup files are saved manually Logger and Wave files are saved automatically Each file consists of some...

Страница 60: ...will be increased by one The number can be changed from 0 to N The only limitation of the N value is the length of the file name which cannot be longer than eight characters In the case when such lim...

Страница 61: ...should be pressed except the Shift or the Alt one Notice The files can be overwritten the use of the same file name without any warning if the Replace option is switched on path Menu File Save Options...

Страница 62: ...f capital letters and have no extensions By pressing the ENTER push button the window with the list of available operations is opening for marked highlighted position ENT When SD Card is selected the...

Страница 63: ...er window is closed and the instrument returns to the File list after pressing the ESC push button 4 4 4 4 To return to the upper catalogue the user has to press the 3 3 3 3 push button The upper high...

Страница 64: ...proposition of new catalogue name will appear at the bottom line of the display If there already exists a directory with such name the warning message will appear In other case this operation will be...

Страница 65: ...files saved in the files memory are moved in order to obtain the continuous occupied space After pressing the ENTER push button on the active Yes option the instrument checks whether the used result a...

Страница 66: ...ration for the catalogue is the same as opening the catalogue by means of the 4 4 4 4 push button ENT Opening the measurement file means that the measurement results saved in this file will be loaded...

Страница 67: ...talogue from the file catalogue list in one memory type to another or from one catalogue of external memory to another catalogue same memory It is not possible to move logger files to the Internal mem...

Страница 68: ...the contents of the memory and make operation on Setup files such as open delete copy move rename files create new files and display file information In the Setup Manager window only setup files are...

Страница 69: ...p push button the instrument will start the measurement with the loaded settings ENT To get information about the setup file one should for selected file press the ENTER push button on the Info positi...

Страница 70: ...he Auto Save function Notice The Auto Save function can be performed only in the case when the Meas Period path Menu Measurement General Settings is not less than 10 seconds If it is less than 10 seco...

Страница 71: ...ing these push buttons the Save window is accessed if the measurements are not performed If the option is active after pressing the ENTER and Alt push buttons the results are saved in the file with th...

Страница 72: ...the Shift and the Start Stop push buttons Multifunction I O it enables the user to select the available functionality of the I O port Power Supply it enables the user to check the powering source of t...

Страница 73: ...conjunction or in a sequence with the Shift one The measurements are started or stopped after pressing both push buttons 8 2 Setting parameters of the I O port Multifunction I O The Multifunction I O...

Страница 74: ...ns Alarm Pulse Polarisation selection of the digital output signal In the Polarisation position it is possible to select which polarisation of the signal negative or positive will be valid Active leve...

Страница 75: ...returns to the Instrument list after pressing the ENTER or ESC push button ENT The time edition is doing in the special window which is opening after pressing the 4 4 4 4 push button The selection of...

Страница 76: ...nsducer s compensation activation Transducers The Transducers position enables the user to activate or deactivate the transducer s noise compensation ENT 8 7 Checking the specification of the instrume...

Страница 77: ...e access to the program functions of the instrument Instrument Mode it enables the user to define the scope of available instrument functions and adjust accordingly the lists in some menu windows Refe...

Страница 78: ...he activation of Partial results in locking access to the Menu options which are responsible for measurement parameters In the case of Full Lock no one position from the Menu lists is accessible and a...

Страница 79: ...SVAN 106 User MANUAL _ 79 Below some windows different in simple and advanced modes are presented 9 8 Reference signal in vibration measurements Reference Levels...

Страница 80: ...ocity position the user can set the reference level of the velocity signal It is possible to set this level from 1 nms 1 to 100 nms 1 9 9 User filter setting User Filters The User Filters position ena...

Страница 81: ...layed after Saving the measurement results in a file When the position is set to be active the special warning can be displayed after pressing the Start Stop push button It will be happened in a case...

Страница 82: ...SVAN 106 User MANUAL _ 82...

Страница 83: ...e frequencies of pass band filters together with three Total Values measured with selected by the user weighting filters are presented in the switched on Spectrum mode 10 1 Selection of 1 1 Octave or...

Страница 84: ...or 1 3 Octave analysis can be saved in the logger file The spectrum saving in the logger file can be activete for each channel in the appropriate window path Menu Measurement Data logging Logger Resu...

Страница 85: ...ters which are mainly dedicated for the control of the spectrum view The following windows contain the elements that influence the presentation of the results of 1 1 Octave and 1 3 Octave analysis ENT...

Страница 86: ...s like below 10 7 Setting the scale of the spectrum results presentation Spectrum Scale The Spectrum Scale sub list enables the user to change the scale in the available modes of graphical presentatio...

Страница 87: ...1 Octave results presented with different Dynamic parameter 80 and 40 dB are presented Switching on off the grid in the graphical mode presentation The Grid enables the user to switch on or off the gr...

Страница 88: ...ndow enables one to select for which channels spectrum will be viewed during multi spectrum display mode and to assign special color for the spectrum curve ENT 10 10 Setting the parameters for total v...

Страница 89: ...10 11 Setting user filter coefficients for 1 1 Octave and 1 3 Octave analysis User Filters The User Filters position path Menu Auxiliary Settings User Filters enables the user to introduce the values...

Страница 90: ...tre frequency filter 100 0dB 100 0dB 0 63 Hz available values for 0 63 Hz centre frequency filter 100 0dB 100 0dB 0 80 Hz available values for 0 8 Hz centre frequency filter 100 0dB 100 0dB 1 00 Hz av...

Страница 91: ...HAV WBV Dosimeter The HAV WBV Dosimeter list is opening from the Measurement menu This list enables one to set up the parameters for dose measurements like exposition period type of measurement whole...

Страница 92: ...set up its own coefficients ENT 11 2 Setting parameters for channels Channel x The Channel x positions enable one to set up or display parameters for the individual channel like input type and filters...

Страница 93: ...ure AEQ MAX RMS EAV Total Time EAV Time Left ELV Total Time ELV Time Left 3 Second field for dose results measurement 4 Elapsed time shows the current second of the measurement The value presented the...

Страница 94: ...is loaded after pressing the ENTER push button The name of this file appears in a list as it is presented below ENT The message Invalid File Content is displayed when the selected file does not includ...

Страница 95: ...results are extrapolated The required value can be set in the special window which is opened by means of the 3 3 3 3 4 4 4 4 push buttons pressed together with the Alt one The Exposure Time can be set...

Страница 96: ...SVAN 106 User MANUAL _ 96 ENT ENT ENT ENT...

Страница 97: ...SVAN 106 User MANUAL _ 97...

Отзывы: