SVAN 106 User MANUAL
_
36
The Logger Setup list enables the user to edit the name of the logger file
(history of results measurements or wave records) and to set other general
parameters. Depending on what Logging Mode was selected (Logger or
Wave) in General Settings window, the Logger Setup window has
different view.
When Logger mode was selected apart from logger file name the user may
define the interval of the data logging in a file (Logger Step). The
Logger Step can be set from 100 milliseconds to 1 hour.
The Logger Name enables the user to
name the logger file. The default one is
LOG# for Logger files and &REC# for
Wave files. The name cannot be longer
than eight characters. After pressing the
<
3
3
3
3
>, <
4
4
4
4
> push-buttons, the special
window with text editor function is opening.
<
4
4
4
4
>
The edited name is accepted and the file is saved after pressing the <ENTER> push-button. The special
warning is displayed in the case the file with the edited name already exists in the memory. The instrument
waits then for a reaction of the user (any push-button should be pressed except the <Shift> or the <Alt>
one).
The main measurement results (cf. App. B) are calculated in the period set on in the Meas. Period. These
results can be saved in the result files in the instrument’s or external memory. In the case the Meas. Period
is greater than 9 seconds, it can be done also by means of the Auto Save operation. In case the
Repetition No. is greater than one, the Auto Save operation will be performed after the period set on in the
Meas. Period. The name of the file with the main results is increased by one after each saving.
In the same, when the Logger Mode is switched on (On), the partial measurement results are calculated in
the period set on in the Logger Step. Up to 60 results can be logged simultaneously from all channels and
profiles of the instrument (PEAK / P–P/ MAX / RMS / VDV) and two vectors (VEC13 and VEC46) with time
step down to 100 ms. These results are saved in one logger file. The name of the file is set on in the
Logger Name position. The registration in the logger’s memory is stopped after the period, which is equal to
Meas. Period multiplied by Repetition No., after pressing the <Start/Stop> push-button or after stopping
the measurements remotely.