![Svantek SVAN 106 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/svantek/svan-106/svan-106_user-manual_1409275044.webp)
SVAN 106 User MANUAL
_
44
The position Enabled switches on/off the Wave Trigger.
In the Rec. Limit position it is possible to select the time of signal recording
after triggering. If the triggering condition appears then the signal will be
recording during the period defined in Rec. Time. The available values of
Rec. Limit are: Max Length, Fixed Len. or Off.
When Off is chosen then the event signal will be recording as long as
memory was filled or until the trigger condition was ceased. When
Max Length is chosen then the signal will be recording during period
defined in the Rec. Time, but can be stopped earlier if trigger condition was
ceased. When Fixed Len. is chosen then the signal will be recording during
period defined in the Rec. Time, even when trigger condition ceased.
In the Rec. Time position it is possible to select the time of event signal
recording after triggering. If next triggering condition appears then the signal
will be recording additional Rec. Time.
<
4
4
4
4
>
The position Trigger Parameters enables
to define the parameters of the triggering
signal. To open this position one should
select it and press <ENTER> push-button.
The Trigger Parameters
window
and
meaning of all positions is identical as for
Logger Trigger case.
<ENT>
5.8 Measure triggering parameters selection – Measure Trigger
The Measure Trigger position appears
only in advanced instrument mode (path:
<Menu>
/
Auxiliary Setup
/
Instrument Mode: Advanced Mode).
The Measure Trigger sub-list enables the
user to set the parameters for triggering the
measurement.
<ENT>
The Measure Trigger is a contexts sub-list in which the triggering can be
switched off or on (Trigger), in the case when on - the source of the
triggering signal can be determined (Source), channel of source signal
(Channel), its level (Level) and sometimes also the speed of changes
(Gradient). If RTC is selected as trigger type, start time (RTC Start) and
repetition of triggering (Repeat Every) is defined.
Switching the triggering on and off