SVAN 106 User MANUAL
_
43
5.7 Wave recording
All positions connected with wave recording
in Data Logging list appear only in
advanced instrument mode (path: <Menu> /
Auxiliary Setup
/
Instrument Mode:
Advanced Mode).
When Wave function is chosen for the
Logging Mode, the Data Logging list
consists
of
three
active
positions:
Logger Setup,
Wave Channels
and
Wave Trigger.
=>
Setting-up the Wave recorder function
The Logger Setup list enables the user to edit the name of the logger file
(history of results measurements or wave records) and to set other general
parameters.
The Wave Rec. position is not active. It only informs about the used type of
wave recording: Continuous.
The user may define the format of the wave file header (Format). Format of
the wave file header may be PCM or Extensible.
The Sampling Rate position is not active. It only informs about the used sampling rate: 6000Hz.
Selecting
the
channels
for
Wave
recording
The Wave Channels position appears only
in advanced instrument mode (path:
<Menu>
/
Auxiliary Setup
/
Instrument Mode: Advanced Mode).
The Channel x positions enable the user to
select the channels which signal will be
recorded.
<ENT>
Wave recorder trigger setup
The Wave Trigger position appears only in
advanced instrument mode (path: <Menu> /
Auxiliary Setup
/
Instrument Mode:
Advanced Mode).
The Wave Trigger enables the user to
activate and programme the wave recorder
trigger.
<ENT>