SVAN 106 User MANUAL
_
80
The Reference Levels sub-list enables the
user to set the reference level of the
vibration signal. The values, which are set
here, are taken into account during the
calculations of the measurement results
expressed in the Logarithmic scale (with
the dB as the units).
<ENT>
In the Acceleration position the user can set the reference level of the acceleration signal from 1 µms
-2
to
100 µms
-2
. In the Velocity position, the user can set the reference level of the velocity signal. It is possible to
set this level from 1 nms
-1
to 100 nms
-1
.
9.9.
User filter setting – User Filters
The User Filters position enables the user to introduce the values of the
coefficients of the user filters. This position is active only in 1/1 Octave and
1/3 Octave modes.
9.10.
Selection of the Vibration units - Vibration Units
The Vibration Units
position enables the
user to select the metric or non-metric
units.
It is possible to select the Non-Metric units
(e.g. g, ips, mil etc.) or Metric units (e.g.
m/s
2
, m/s, m etc.).
<ENT>
9.11.
Warnings selection – Warnings