Additionally to the bandpass RMS results three Total values are measured during 1/1 and 1/3 octave
analysis. Total values parameters (weighting filter, type of integration filer for acceleration, velocity or
displacement results and additional calibration factor) are setting up in the Total Values window (path:
<Menu> / Display / Spectrum Display Setup).
The output of the selected 1/1 Octave or 1/3 Octave filter can be also used as the triggering signal for
different trigger applications: triggering the logger, triggering the event recording, triggering the wave
recording and triggering the alarms.
10.3.
Activation of saving of 1/1 Octave and 1/3 Octave analysis results in the
logger’s file - Logger Results
The RMS results from 1/1 Octave or
1/3 Octave analysis can be saved in the
logger file. The spectrum saving in the
logger file can be activete for each channel
in the appropriate window (path: <Menu> /
Measurement
/
Data logging
/
Logger Results / 1/1 Octave or 1/3 Octave
).
<ENT>
If the None value is selected in the position Channel x then spectrum data
will not be saved in the logger file. If RMS value is selected, then RMS
spectrum will be saved in the logger file for this channel.
10.4.
Selection of the 1/1 Octave and 1/3 Octave bandpass results as triggering
source
For the 1/1 Octave or 1/3 Octave analysis functions it is possible to define
trigger events for logger, event, wave and alarm triggers, based on the
spectrum selected band level.
This trigger condition for above applications can be programmed in
windows: Logger Trigger, Event Recording Trigger, Wave Trigger and
Alarm Trigger. In these lists one should set up the Source to Spectrum
and select required band pass in the Result position.