89
SV
104BIS User Manual
•
Successful firmware update will be indicated by relevant message.
•
Turn off the unit.
Note:
With use of
Supervisor
software it is very easy to check if there are any new firmware
releases available for download (see
7.7
S
TORING THE INSTRUMENT
•
To preserve the life of the internal batteries, it is recommended that the SV 104BIS instrument
is turned off when it is stored.
•
It is best not to store the dosimeter for any period of time in low battery state condition.
•
Generally, the SV 104BIS instrument should be stored at room temperature, charged to about
40 to 60% of capacity.
•
After use, it is best to charge SV 104BIS prior to storage for a prolonged period that is longer
than 3 months.
•
It is recommended to turn off the Bluetooth
®
interface in applied settings.
•
If the dosimeter is to be stored for an even longer period, it is best to take the instrument out of
storage and re-charge the dosimeter every 8 months. When SV 104BIS is turned off, it still
draws a small amount of battery power, so periodic recharge is to prevent battery over-
discharge.
7.8
T
RANSPORTATION AND CARRYING
For transportation or storage purpose, always use the packaging provided by the manufacturer. In a potentially
dirty industrial environment, it is advisable to use the carrying case provided by the manufacturer
SA 144
,
which ensures excellent mechanical and environmental protection and long-term storage conditions.
The temperature range over which the device can be stored/transported is -
20°C to +50°C
.
Note:
For Air-transport Bluetooth
®
should be
turned off !
7.9
T
ROUBLESHOOTING
1. Upon connection onto the docking station port, if automatic charging is not started: check the docking
station LED colour for state of charging (Chapter
2. If the incorrect time or date is displayed when turning on the instrument connect the device to the
computer and use
Supervisor
software to set the time and date (see) ensuring PC clock is set
correctly.
3. In case the instrument is not able to turn on ensure the unit is charged by connecting onto the docking
station. This ensures the battery is not exhausted. Then proceed with hardware reset (Chapter
4. In case your dosimeter does not respond proceed with turn-off/turn-on procedure (Chapter
),
and hardware reset of the instrument (Chapter
).
5. In case the measurement of the sound level is frozen or set to a fixed value proceed with turn-off/turn-
), then with hardware reset of the instrument (Chapter
6. In case the reset does not help proceed to Chapter
Содержание SV 104BIS
Страница 139: ...139 SV104BIS User Manual 250 Hz 315 Hz 400 Hz 500 Hz...
Страница 140: ...140 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 141: ...141 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 142: ...142 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 143: ...143 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 144: ...144 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 148: ...148 SV104BIS User Manual Figure C 4 Total directional characteristics asymmetrical axis...
Страница 150: ...150 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 151: ...151 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 152: ...152 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 153: ...153 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 154: ...154 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 161: ...161 SV104BIS User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 162: ...162 SV104BIS User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 166: ...166 SV104BIS User Manual 1 3 octave filters for 16 0 Hz 1 1 octave filter...
Страница 180: ...180 SV104BIS User Manual...