93
SV
104BIS User Manual
#1;
The instrument outputs in this case a file containing all control settings given in Tab. A1 in the format:
#1,Xccc,Xccc,(...),Xccc;
Example:
The instrument sends the following sequence of characters as an answer for the mentioned above
request:
#1,U104,N12342,W1.08.1,Q0.01,M4,F2:1,F3:2,F1:3,J2:1,J3:2,J1:3,f1,C1:1,C0:2,C2:3,B0:1,B3:2,B15:3,
b0,d1s,D10s,K5,L0:1,L0:2,L0:3,Y3,y0,XC115:1,XC115:2,XC115:3,Xl115:1,Xl115:2,Xl115:3,S0,T1,e480
,c1:1,c1:2,c1:3,h0:1,h0:2,h0:3,x3:1,x3:2,x5:3;
means that:
•
SV 104BIS is investigated (U104);
•
number is 12342 (N12342);
•
software version number is 1.08.1 (W1.08.1);
•
calibration factor is equal to 0.01 dB (Q0.01);
•
DOSE METER
is selected as the measurement function (M4);
•
A
filter is selected in profile 1 (F2:1);
•
C
filter is selected in profile 2 (F3:2);
•
Z
filter is chosen in profile 3 (F1:3);
•
A
Peak filter is selected in profile 1 (J2:1);
•
C
Peak filter is selected in profile 2 (J3:2);
•
Z
Peak filter is selected in profile 3 (J1:3);
•
Z
filter is selected for
1/1 OCTAVE or 1/3 OCTAVE
analysis (f1)
•
FAST
detector is selected in profile 1 (C1:1);
•
IMPULSE
detector is chosen in profile 2 (C0:2);
•
SLOW
detector is selected in profile 3 (C2:3);
•
logger’s buffer is not filled by the results from profile 1 (B0:1);
•
PEAK
and
MAX
values are stored in the files of the logger from profile 2 (B3:2);
•
PEAK
,
MAX
,
MIN
and
LEQ
values are stored in the files of the logger from profile 3 (B15:3);
•
results of
1/1 OCTAVE
analysis are not stored in the files of the logger (b0);
•
results are stored in a logger’s file every 1
second (d1s);
•
integration period is equal to 10 seconds (D10s);
•
measurement has to be repeated 5 times (K5);
•
linear detector is selected to the
LEQ
calculations in profile 1 (L0:1);
•
linear detector is selected to the
LEQ
calculations in profile 2 (L0:2);
•
linear detector is selected to the
LEQ
calculations in profile 3 (L0:3);
•
delay of the start of the measurements is equal to 3 seconds (Y3);
•
synchronization the start of measurement with RTC is switched off (y0);
•
threshold level for PTC calculation in profile 1, is set to 115 dB (XC115:1);
•
threshold level for PTC calculation in profile 2, is set to 115 dB (XC115:2);
•
threshold level for PTC calculation in profile 3, is set to 115 dB (XC115:3);
•
threshold level for ULT calculation in profile 1, is set to 115 dB (Xl115:1);
•
threshold level for ULT calculation in profile 2, is set to 115 dB (Xl115:2);
•
threshold level for ULT calculation in profile 3, is set to 115 dB (Xl115:3);
•
instrument is in the Stop state (S0);
•
logger is active (T1);
•
exposition time is set to 8 hours (e480);
•
criterion level in profile 1 is chosen as 80 dB (c1:1);
Содержание SV 104BIS
Страница 139: ...139 SV104BIS User Manual 250 Hz 315 Hz 400 Hz 500 Hz...
Страница 140: ...140 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 141: ...141 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 142: ...142 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 143: ...143 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 144: ...144 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 148: ...148 SV104BIS User Manual Figure C 4 Total directional characteristics asymmetrical axis...
Страница 150: ...150 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 151: ...151 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 152: ...152 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 153: ...153 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 154: ...154 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 161: ...161 SV104BIS User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 162: ...162 SV104BIS User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 166: ...166 SV104BIS User Manual 1 3 octave filters for 16 0 Hz 1 1 octave filter...
Страница 180: ...180 SV104BIS User Manual...