Note:
It is advised to perform an acoustic calibration of the instrument each time before and after
the measurement run. A single calibration at the start of each day of use is usually sufficient for
most regulations.
Note:
The calibration factor is always added to the results in the Dosimeter or 1/1-octave or
1/3-octave analyses modes.
Note:
The manufacturer’s recommended factory calibration interval is every 12 months for this
instrument to be confident in its continuing accuracy and compliance with the international
specifications. Please contact your local SVANTEK distributor for further details.
To calibrate the instrument manually the user has to enter the
Calibration
menu.
1. Adjust the actual calibration level of the calibrator to be used with the
and
>
keys.
Figure 4-8
Calibration menu
2. Attach the acoustic calibrator
SV 34B
(or equivalent 114dB@1000Hz) carefully yet tightly over
the microphone of the instrument.
Note:
It is also possible to use an electro-mechanical pistonphone, which generates the signal
(ca
124 dB) or different type of acoustic calibrator dedicated for ½” microphones with an alternative
output level such as 94 dB at 1 kHz. In any case, before starting the calibration measurement, the
user has to set the level of the signal generated by the given calibrator (Level position), which is
stated in the calibration certificate of the unit (the default expected value of the Calibration Level set
by the manufacturer of SV 104BIS
is equal to 114 dB).
3. Switch on the calibrator and wait ca 30 seconds for the tone to stabilise before starting the calibration
measurement.
4. Start the calibration measurement by pressing the
key.
5.
The calibration measurement time is set to 1 second with 3 seconds’ delay. Calibration stops when
either 5 consecutive results will not differ from themselves by more than 0.02 dB or 10 consecutive
results will not differ from themselves by more than 0.05 dB. It is possible to stop the calibration
measurement by pressing two keys at the same time
and
.
Figure 4-9
Calibration - initial delay screen
6. Delay before starting calibration measurement is counted down on the display. After the measurement,
its result is presented on the display.
Содержание SV 104BIS
Страница 139: ...139 SV104BIS User Manual 250 Hz 315 Hz 400 Hz 500 Hz...
Страница 140: ...140 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 141: ...141 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 142: ...142 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 143: ...143 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 144: ...144 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 148: ...148 SV104BIS User Manual Figure C 4 Total directional characteristics asymmetrical axis...
Страница 150: ...150 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 151: ...151 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 152: ...152 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 153: ...153 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 154: ...154 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 161: ...161 SV104BIS User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 162: ...162 SV104BIS User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 166: ...166 SV104BIS User Manual 1 3 octave filters for 16 0 Hz 1 1 octave filter...
Страница 180: ...180 SV104BIS User Manual...