104BIS User Manual
Measured
quantities
The measured quantities for
DOSE METER
mode are:
Time, Lpeak, Lmax, Lmin,
SPL (
L
)
, DOSE, D_8h,
PrDOSE, Lav, Leq,
SEL
(
LE
)
, SEL8, E, E_8h, LEPd, PSEL, Ltm3, Ltm5,
Leq statistics
(
Ln
)
, PTC, PTP, ULT,
TWA, PrTWA, Lc-a.
Definitions for measured quantities are given in Appendix D.
Conformance
testing
This chapter contains the information needed to conduct conformance testing according to the specified
standards.
Mounting for acoustical tests
The microphone must be mounted on the instrument.
Electrical substitute for the microphone
To obtain a BNC Class electrical input, the microphone
must be replaced by an electrical microphone impedance
adapter SC 104AT
before turning the instrument on
. Total
microphone substitute impedance is 125
Ω.
Note:
The recommended time interval for periodic test of noise exposure meter for checking its
acoustic and electrical working characteristics is one year.
Note:
For the conformance electrical tests the
Microphone Compensation
must be disabled (see
Chapter
Note:
For the conformance acoustical tests with the microphone the
Microphone Compensation
For the comparison coupler evaluation, the
Free Field
compensation must be disabled.
For the free filed evaluation, the
Free Filed
compensation must be enabled.
Linear
Operating
Ranges
The starting point at which tests of level linearity shall begin is 114.0 dB for the frequencies specifies below.
Linear operating
range for the sinusoidal signal and microphone sensitivity 1 mV/Pa
Single measuring range
–
see table below.
Table C.1
Linear operating ranges for Leq
[dB]
L
AS/F
L
CS/F
L
ZS/F
L
AeqT
L
CeqT
L
AE
L
Cpeak
(t
int
= 2 s)
from
to
from
to
from
to
from
to
from
to
from
to
from
to
31.5 Hz
53
100.6
55
137
65
138
53
100.6
53
137
53
103.6
56
140
500
Hz
53
136.8
55
140
65
138
53
136.8
53
140
53
139.8
56
143
1 kHz
53
140
55
140
65
138
53
140
53
140
53
143
56
143
4 kHz
53
141
55
139.2
65
138
53
141
53
139.2
53
144
56
142.2
8 kHz
53
138.9
55
137
65
138
53
138.9
53
137
53
141.9
56
141
Note:
For the signals with the crest factor
n >1.41
upper measuring range of the RMS (
LEQ
and
SPL
) is reduced. The valid upper limit can be calculated according to the below given formula:
𝑨
𝒏
= 𝟏𝟒𝟎 − 𝟐𝟎 𝒍𝒐𝒈(𝒏/√𝟐)
, where
A
is the upper limit for the sinusoidal signal
Example:
For the crest factor
n = 10
the upper limit is
A10
=
123 dB
Содержание SV 104BIS
Страница 139: ...139 SV104BIS User Manual 250 Hz 315 Hz 400 Hz 500 Hz...
Страница 140: ...140 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 141: ...141 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 142: ...142 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 143: ...143 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 144: ...144 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 148: ...148 SV104BIS User Manual Figure C 4 Total directional characteristics asymmetrical axis...
Страница 150: ...150 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Страница 151: ...151 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 152: ...152 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 153: ...153 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Страница 154: ...154 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 161: ...161 SV104BIS User Manual 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 162: ...162 SV104BIS User Manual 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter 31 5 Hz 1 1 octave filter...
Страница 166: ...166 SV104BIS User Manual 1 3 octave filters for 16 0 Hz 1 1 octave filter...
Страница 180: ...180 SV104BIS User Manual...