![Suez Sievers UPW Boron Analyzer Скачать руководство пользователя страница 92](http://html1.mh-extra.com/html/suez/sievers-upw-boron-analyzer/sievers-upw-boron-analyzer_operation-and-maintenance-manual_1389722092.webp)
SUEZ © 2018
DLM 71000-03 EN
Rev. B
Chapter 8: Maintenance
Formatting a New Data Card
When the Analyzer powers up, it will display a message about the data card not being formatted or that the
data card is from another Analyzer. A "C" will also appear in the maintenance alert area of the screen.
1. Press
SETUP
from the
Main Menu
to access the
SETUP
menu.
2. Highlight
MAINTENANCE
and press
ENTER.
3. Highlight
DATA CARD
and press
ENTER
.
4. Press the
FRMAT
key to format the data card.
5. Press
YES
to the question, “Do you wish to continue?” The Analyzer formats the data card.
6. Press
NEXT
to view the size of the card and the days remaining.
7. Press
EXIT
three times to return to the
Main Menu
.
Verification and Calibration
Users should not need to perform either verification or calibration as part of normal operation or routine
maintenance. The Analyzer is initially calibrated and verified at the factory using highly reliable conductivity
cells. This calibration is stable and does not require checking at regular intervals.
Cleaning Instructions
When cleaning the instrument, use only a clean cloth dampened with water.
Returning Product to SUEZ
Save the shipping carton and packing materials that came with the Analyzer. If for any reason the Analyzer
must be returned to the factory, it must be packed in the original carton to ensure that no damage occurs
during shipment. There is a charge for new packaging materials to return the Analyzer. The Analyzer must
be insured during all phases of shipment.
Prior to shipment it is critical that all water be purged from the Analyzer's system:
1. Shut off water to the Analyzer and disconnect the inlet lines.
2. Run the Analyzer in the
STANDBY
mode to empty the excess water.
Note:
The calibration of the Analyzer should not change. If there is a significant discrepancy
between the Expected Boron Value and Measured Boron Value reported at the end of the
verification process, contact Technical Support or your local service provider.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...