Chapter 5: System Configuration
Figure 10: Front Panel Display Showing the Main Menu in the Stopped State
START
Press to begin the mode of operation displayed.
VIEW
Press to display the six available options:
BORON, CAL/VERIFY HIST, MAINTENANCE, SYSTEM
CONFIG, MESSAGES,
and
DEBUG.
BORON
The BORON option is used to view the Boron Data. To access the BORON screen:
1. Press View from the
Main Menu
to enter the
View
menu.
2. Use the
UP
and
DOWN
arrow keys to highlight
BORON
.
3. Press
ENTER
to display stream number options.
4. Highlight the appropriate stream and press
ENTER
to display the options for each stream.
5. Press
ENTER
while the
BORON DATA
option is highlighted to display the most recent data for the
stream.
• Up to twenty-four historical values are displayed as text.
• The data is displayed with appropriate units and time measured.
• The date when the data was measured is in the upper-right corner.
• Three significant figures are displayed.
• Use the
UP
and
DOWN
arrow keys to advance through measurements.
• The most recent data is displayed in the upper-left position of the values displayed.
• If the
UP
or
DOWN
arrow key is held, data advances one page at a time.
• If the data is viewed during analysis, a newly acquired data point will update the screen at the
upper-left position.
• The date of the measurement will also be replaced with the countdown time left in analysis in
the upper-right corner.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...