Chapter 6: Setting Up Inputs and Outputs
digital to analog setting associated with the 20 mA signal is highlighted to show that it can be edited. Using
the arrow keys, enter the boron level that will correspond to the 20 mA output, and press
NEXT
to save the
digital-to-analog settings and return to the previous menu.
To set the analog output to
LOW OUTPUT
, scroll to
LOW OUPUT
and press
NEXT
. This setting will make
the selected analog output yield a constant 4 mA signal. When the analog output setting for each analog
output has been set, press
EXIT
four times to return to the
Main Menu
.
To set the level for the analog outputs when the system is in standby mode, highlight
ANALOG STANDBY
LEVEL
and press
ENTER
. Using the arrow keys, select the desired analog level; options are
0 mA
,
2 mA
and
4 mA
.
Alarm Output
The alarm output is a signal that is sent from the Analyzer when predetermined conditions are met within
the Analyzer. There are ten choices available for an alarm output:
• OFF
• BORON STREAM X
• STATE
• RES ERROR ANY STREAM
• ERROR
• RES ERROR STREAM X
• READY STREAM X
• RES WARN ANY STREAM
• BORON ANY STREAM
• RES WARN STREAM X
Use the
UP
and
DOWN
arrow keys to scroll through the ten settings. The
OFF
setting disables the alarm
output and is the default setting for both of the alarm outputs. The
STATE
setting will send an alarm signal
when the Analyzer is stopped (either by normal means or a fatal error). The
ERROR
setting enables the
alarm output to send an alarm signal when an error condition occurs on the Analyzer. The
BORON ANY
STREAM
and
BORON STREAM X
settings enable the alarm output to send an alarm signal when a preset
threshold is met. The
RES ERROR
and
RES WARN
settings enable the alarm output to signal an alarm
when the Resistivity is low enough to trigger an error or warning. The
READY STREAM
X enables the
alarm output to signal an alarm when a boron measurement has completed on the specified stream.
Note:
Pressing
EXIT
instead of
NEXT
will exit the analog output editing menu without
saving any changes to the 4 mA and 20 mA values.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...