Chapter 4: Basic Analyzer Operation
Figure 6: Fluidics Block Diagram
Zero Measurement
During a zero measurement, the UPW sample is diverted through the zero column to remove any boron
before it is sent to the injection valve. The zero is measured as the conductivity peak when an injection of
reagent is made into a UPW sample that is free of boron.
Sample Measurement
During a sample measurement, the UPW sample is sent directly to the injection valve. The sample is
measured as the conductivity peak when an injection of reagent is made into a UPW sample containing
boron. The boron concentration is proportional to the difference between the sample and zero conductivity
peaks.
The number of sample measurements selected applies to all streams. After the designated number of
samples has been analyzed, the Analyzer advances to the next available stream, zeros itself, and begins
analysis.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...