Chapter 5: System Configuration
• To view older data, use the
RIGHT
arrow key to display the date and type of calibration/
verification of older records.
• The codes indicating the calibration/verification type are as follows:
CCS
: Cell Slope Calibration
CCZ
: Cell Offset Calibration
VER
: Verification
3. Press
EXIT
at any point in the
CAL/VERIFY HIST
menus to return to the previous menu or twice to
return to the
Main Menu
.
MAINTENANCE
The
MAINTENANCE
option allows you to view the maintenance status of the five maintenance items.
1. Press
VIEW
from the
Main Menu
to view the status.
2. Highlight
MAINTENANCE
by using the arrow keys and press
ENTER
. Pressing
ENTER
with any
option highlighted will display the maintenance status of that item.
• For the
PUMP TUBING
,
REAGENT
,
ZERO BORON RESIN
, and
DI RESIN
maintenance items,
the age of the item is displayed along with the time remaining until the component must be
replaced.
• For the
DATA CARD
maintenance item, the size of the data card is displayed in megabytes,
with the time remaining before the data card becomes completely full displayed in days.
3. Press
EXIT
once to return to the previous menu or three times to return to the
Main Menu
. Using
the
VIEW
menu only allows for viewing the status of any of the five maintenance items.
SYSTEM CONFIG
The
SYSTEM CONFIG
option displays the Analyzer's serial number and firmware revision levels.
1. Press
VIEW
from the
Main Menu
to view the system configuration.
2. Highlight
SYSTEM CONFIG
by using the arrow keys. Pressing
ENTER
with
SYSTEM CONFIG
option highlighted will display the Analyzers serial number and firmware revision levels.
3. Press
EXIT
once to return to the previous menu or twice to return to the
Main Menu
.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...