Chapter 5: System Configuration
When the stream name is displayed or printed, two characters ("-x") are appended to the 12-character
stream name, with "x" indicating stream number 1, 2, 3, or 4.
To select
STREAM ENABLED/DISABLED
:
1. Highlight that field using the
UP
and
DOWN
arrow keys and press
EDIT
.
2. Use the
UP
and
DOWN
arrow keys to toggle between
ENABLED
or
DISABLED
.
3. Press
ENTER
to save the change and return to the
STREAM SETTINGS
menu.
To edit the number of
SAMPLES/STREAM
:
1. Highlight that field using the
UP
and
DOWN
arrow keys and press
EDIT
.
• If the
SAMPLE MODE
has been set to
INDIVIDUAL
, the display will show the number of
individual samples that will be reported for that stream.
• If the
SAMPLE MODE
has been set to
AVERAGED
, the display will show the number of
averaged results that will be reported for that stream. Each averaged result consists of
5 sample measurements that are averaged after the fifth measurement is complete. For
example, if the display shows "# Averaged Measurements: 4", this indicates that 20 samples
will be measured for this stream, with 4 averaged results being reported (one after every fifth
sample measurement).
2. Use the
UP
and
DOWN
arrow keys to increase or decrease the number of samples.
• In
INDIVIDUAL
mode, the number will increase from 5 to 240 in increments of 5.
• In
AVERAGED
mode, the number will increase from 1 to 48 in increments of 1.
3. Press
ENTER
to save the change and return to the
STREAM SETTINGS
menu. The bottom of the
display will show the updated amount of time the Analyzer will spend analyzing that stream before
advancing to the next stream.
ADVANCED
Selecting the
ADVANCED
option from the
Setup
menu opens a submenu that allows you to access
DIAGNOSTICS, CAL. CONSTANTS, CELL CONSTANTS,
and
SERIAL PORT
. From the Setup menu:
1. Highlight
ADVANCED
and press
ENTER
.
DIAGNOSTICS
This options runs a series of tests used to determine if any of the electronics are performing outside of their
specifications. Performing diagnostic testing should only be done in the event of an electronics malfunction
detected by the firmware and you have been directed to perform such testing by Technical Support or your
local service provider. To access the DIAGNOSTICS menu:
1. Press
SETUP
from the
Main Menu
, highlight the
ADVANCED
menu option.
2. Select
DIAGNOSTICS
, and press
ENTER
. For information on diagnostic testing, contact Technical
Support or your local service provider.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...