![Suez Sievers UPW Boron Analyzer Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/suez/sievers-upw-boron-analyzer/sievers-upw-boron-analyzer_operation-and-maintenance-manual_1389722066.webp)
SUEZ © 2018
DLM 71000-03 EN
Rev. B
Chapter 6: Setting Up Inputs and Outputs
When permanently wiring to any of the I/O ports, it is important to install the supplied liquid-tight fittings in
the wiring knockouts on the left side of the Analyzer. Be sure to run the cable through the fitting before
connecting it to the terminal block.
Analog Outputs
The analog output is the way in which data is translated into a signal for reading at another location. The
analog outputs are isolated from the main circuit. To set up the analog outputs, proceed as follows:
Step One: Wire Terminal Block TB6
The data from the Analyzer may be recorded using the 4-20 mA outputs located on terminal block TB6.
The Analyzer has four analog outputs located on the terminal block. The analog outputs are calibrated
prior to shipping.
Table 2: Analog Outputs on Terminal Block TB6
Step Two: Configure the Analyzer
Once the terminal block is wired, wire the next input/output or replace the card cage cover and close the
Analyzer door, turn on the Analyzer, and proceed as follows:
1. Press
SETUP
from the
Main Menu
to access the
SETUP
menu.
2. Using the
UP
and
DOWN
arrow keys, highlight
INPUT/OUTPUT
and press
ENTER
.
3. Highlight
ANALOG OUTPUTS
and press
ENTER
. The next menu displays the options
ANALOG
OUTPUTS
and
ANALOG STANDBY LEVEL
.
Содержание Sievers UPW Boron Analyzer
Страница 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 24: ...SUEZ 2018 24 of 104 DLM 71000 03 EN Rev B Chapter 1 Introduction THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 64: ...SUEZ 2018 64 of 104 DLM 71000 03 EN Rev B Chapter 5 System Configuration THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 96: ...SUEZ 2018 96 of 104 DLM 71000 03 EN Rev B Chapter 8 Maintenance THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 100: ...SUEZ 2018 100 of 104 DLM 71000 03 EN Rev B Chapter 9 Message and Error Codes THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Страница 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...