![Stryker MMP200 Скачать руководство пользователя страница 87](http://html.mh-extra.com/html/stryker/mmp200/mmp200_installation-and-service-manual_1385968087.webp)
87
S
14.3 Standard Width Yoke
Loosen
the screw
Yoke
Apparatus
Yoke
Handle
1. If yoke does not have a handle, follow the
steps below:
a. Loosen the middle screw on the yoke
apparatus; do not remove the screw
completely.
b. Insert the handle and align the groove
on the handle with the screw hole on the
yoke apparatus.
c. Tighten the screw to secure the handle
in place.
2. Remove the shroud cover by removing the
two knurled screws.
3. Install the Flat Panel Monitor on the Flat
Panel yoke and complete all video and
power connections.
4. Place the power brick in the special carrier.
5. Place cables in cable holders on inside of
the Flat Panel Yoke so that no cables can
be seen from the front or side. Secure the
power cable in a similar fashion.
6. Place the shroud cover and the two knurled screws back into their original seating.
14.4 Halogen Light Flat Panel Stop Replacement
Light Flat Panel Stop Kit (0682-400-163)
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...
Страница 149: ...Stryker Communications 1410 Lakeside Pkwy Flower Mound TX 75028 t 972 410 7100 www stryker com ...