71
S
6. Connect the tubing for supplying the brakes to the pressure regulator mounted above the Ser-
vice Head.
7. Adjust the regulator for 80 (±5.0) PSIG.
8. Test the brakes:
•
Verify the arms remain in place with moderate force applied.
•
Press and hold down both brake buttons.
You should hear a short hissing sound as the brake bladders bleed off. Ensure the Booms can
be easily moved.
•
Individually check each arm by pressing one button at a time.
Note
If air seeps continuously when a button is pressed, the lines connected to it may
be reversed. The button should connect the source of air to the bladder when not
pressed and shut off the air supply on Port One of the valve while venting Port
Two to atmosphere.