![Stryker MMP200 Скачать руководство пользователя страница 75](http://html.mh-extra.com/html/stryker/mmp200/mmp200_installation-and-service-manual_1385968075.webp)
75
S
12.4 Cable Covers
12.4.1 Flat Panel/Light/Light Suspension
Cables
Rubber grommet
Flat Panel Monitor Cables
with Grommet and Cable Tie
1. Route the Flat Panel Monitor cables
through the one-inch opening in the down
tube and attach the down tube to the sus-
pension.
2. Rotate the upper extension arm in both
directions to ensure that sufficient cable
remains outside the down tube.
The arm should be able to meet both limit
stops.
3. Pull the cables out of the down tube an
additional inch to ensure sufficient cable is
available.
4. Wrap a split rubber grommet around the
cables and secure the grommet with a cable
tie.
The rubber grommet prevents the cables
from re-entering the down tube.
Assembled Cable
Cover
Assembled Cable Cover
5. Place the two halves of the Rotating Cable
Cover around the down tube to conceal the
cables.
6. Use two 6-sheet metal screws to attach the
halves together.
7. Use four 6-32 screws to attach the cover to
the base.
12.4.2 Light/Flat Panel Suspension - Teacup Installation
For SHA suspensions, fasten the dual suspension Flat Panel cable cover to the lowest Extension Arm
with four M3x8 screws.
For EHA suspensions, fasten the dual suspension Flat Panel cable cover to the lowest Extension Arm
with two M8x16 screws.
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...
Страница 149: ...Stryker Communications 1410 Lakeside Pkwy Flower Mound TX 75028 t 972 410 7100 www stryker com ...