![Stryker MMP200 Скачать руководство пользователя страница 133](http://html.mh-extra.com/html/stryker/mmp200/mmp200_installation-and-service-manual_1385968133.webp)
133
S
7. Reinstall any vertical separators or UL brackets into the extrusions (mentioned in in Section
17.5.3, Removing Extrusions).
8. Tighten all screws in the end modules.
17.8 Replacing the Motor
WARNING
Power should be cut
from the EP module,
Motor and other elec-
trical modules of the
service head prior to
removing any modules.
1. Dismantle the motor cover halves.
Terminal
Block
2. Remove the purple, blue, and brown wires.
3. Remove the cable clip at the capacitor. The
motor connection cable with now hang
loose.
4. Loosen three of the four Allen screws from
the top (shown in yellow) with a 4mm Al-
len wrench.
Caution
Do not lose the re-
moved Allen screw.
5. Loosen the side Allen screw.
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...
Страница 149: ...Stryker Communications 1410 Lakeside Pkwy Flower Mound TX 75028 t 972 410 7100 www stryker com ...