10.2 Adjusting the Yoke
10.2.1 Adjusting the Width
To adjust the width of the yoke:
1. Remove the power supply from the Central Monitor Support Bracket.
2. Remove the cable cover that is concealing the cables on the left Pivot Arm.
Width Adjust Clamp
Central Monitor
Support
Left Pivot Arm
Cable Cover
Right Pivot Arm
Cable Cover
Remove the cable cover
Note
It is not necessary to remove the cable cover from both Pivot Arms, only the left
side where the cables are concealed.
3. Remove the Central Support Bracket by removing the eight M6 socket head screws and all
washers using a 5mm Allen wrench.
Caution
Before beginning to remove the screws holding the Width Adjust Clamp
together, loosely place two cable ties around each side of the clamp to pre-
vent the yoke from falling once the screws have been removed.