137
S
8. Remove protective caps and four screws that attached shelf to
service head.
9. Remove shelf and pull air lines out of the back of the shelf.
10. If a StrykeVac is installed, pull red wire through new shelf
and attach shelf.
Black Wire
Red Wire
11. Place wires in connector and connect Strykvac.
12. Reattach Strykvac panel.
13. Pull air line through new shelf.
14. Reattach shelf and replace plastic covers.
Supply or Shelf
(if used)
To Brake
15. Insert the supply tubing into the Center Solenoid Connector.
16. Insert the Brake Bladder tubing in the outboard Solenoid
Connector.
17. Turn on the air supply and power to boom.
18. Verify that no leaks are heard and all power to boom works.
Ceiling Brake
19. Verify the operation matches the label on the rail to the tub-
ing coming from the green port on the matching solenoid of
the shelf brake.
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...
Страница 149: ...Stryker Communications 1410 Lakeside Pkwy Flower Mound TX 75028 t 972 410 7100 www stryker com ...