17.7 Reassembling the Service Head
Caution
Power should be cut from the boom system prior to removing any mod-
ules.
17.7.1 Removing and Attaching the Front and Back Plates
1
2
3
4
5
6
To remove the front and back plates:
1. Use a Phillips head screw driver to loosen the eight M4x12 screws (Item 2 in the figure above)
holding the plate to the Flexis Service Head. Do not remove the screws completely, as they are
captive and should remain attached to the face plate.
2. Note the tether (Item 5) attached to the face plate (Item 1) and top plate (Item 6). Do not re-
move the tether, to prevent damage to ground wire.
To attach the front and back plates:
1. Align the face plate so that it is flush and fitted into the open service head.
2. Attach the face plate by loosely threading the eight M4x12 screws.
3. Tighten all screws completely.