![Stryker MMP200 Скачать руководство пользователя страница 139](http://html.mh-extra.com/html/stryker/mmp200/mmp200_installation-and-service-manual_1385968139.webp)
139
S
11. Screw the button onto the Solenoid.
Adjustment Nut
12. Adjust the lock nut from behind to tighten the assembly.
13. Repeat the procedure for the second button.
Supply or Shelf
(if used)
To Brake
14. It there is only one button in the Service Head, insert the sup-
ply tubing into the Center Solenoid Connection.
15. Insert the Brake Bladder tubing in the outboard Solenoid
Connector.
16. Turn on the air supply.
17. Verify that no leaks can be heard.
Ceiling Brake
Mid Brake
FLEXiS Brake Buttons
18. Verify the operation matches the label on the rail to the tubing coming from the green port on
matching the Solenoid of the shelf brake.
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...