S
Description of Control Board
Stryker Visum Electronic
Control System
Stryker Visum Wall
Control
Micro-controller
Siemens SAF- C515 - C
Siemens SAF- C515 - C
Clock rate
8 MHz
8 MHz
Program memory
EPROM , capacity 256 Kbit to 1
Mbit
Flash EPROM, capacity 256
Kbit to 1 Mbit
Parameter memory
EEPROM, capacity 1 KB
N/A
Supply voltage
16 .. 40 VDC ripple max.10%
16 .. 40 VDC +- 10%
Power consumption in a typical
operating situation without exter-
nal components
7 Watt at 40 VDC supply
1 W at 24 V
Current in the light branch
max. 16 A
-
Optional output High side
24V max. 100mA short circuit
proof
-
Optional input
24V pulling against mass
-
Overlap time in case of power
failure
min. 500 ms
-
Printed circuit board
100mm X 160mm
4 - layers multilayer
45mm X 90mm
4 - layers multilayer
Operating temperature (T
A
)
0 .. 85°C
0 .. 70°C
Storage temperature
-40°C .. + 85°C
-40°C .. + 85°C
Relative humidity
Max. 95% at 25˚ C,
no exposure to moisture
Max. 95% at 25°C,
no exposure to moisture
EMC inspection for control and operating console
Interference emission
EN 55011 / 3.91 or
N/A
Interference emission
EN 55011 / 3.91 or
EN 55022 / 8.94
EN 50081-1 / 1.92
N/A
Interference immunity
EN 50082-2 / 3.95
Supply 2KV
N/A
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...
Страница 149: ...Stryker Communications 1410 Lakeside Pkwy Flower Mound TX 75028 t 972 410 7100 www stryker com ...